Paradoxical
enhancement - the unpredicted change of signal intensity of vessels with flowing
blood in multislice imaging. Paramagnetic
- a substance with a small but positive magnetic susceptibility (magnetizability).
The addition of a small amount of paramagnetic substance may greatly reduce the
relaxation times of water. Typical paramagnetic substances usually possess an
unpaired electron and include atoms or ions of transition elements, rare earth
elements, some metals, and some molecules including molecular oxygen and free
radicals. Paramagnetic gadolinium chelate complexes are used as T1 contrast agents
in MRI. Partial
saturation (PS) - excitation technique applying repeated RF pulses in times
on the order of or shorter than T1. In magnetic resonance imaging systems, although
it results in decreased signal amplitude, there is the possibility of generating
images with increased contrast between regions with different relaxation times.
Produces proton density- and T1-weighted images. The change in NMR signal from
a region resulting from a change in the interpulse time, TR, can be used to calculate
T1 for the region. Although partial saturation is also commonly referred to as
saturation recovery, that term should properly be reserved for the particular
case of partial saturation in which recovery after each excitation effectively
takes place from true saturation. PS (SR) images usually have poor image contrast. Permanent
magnet - a magnet whose magnetic field originates from permanently magnetized
material. Permeability
(mu) - tendency of a substance to concentrate magnetic field mu = B/H. Phantom
- an artificial object of known dimensions and properties used to test aspects
of an imaging machine. Phase
- in a periodic function (such as rotational or sinusoidal motion), the position
relative to a particular part of the cycle. Phase
encoding - the application of a gradient prior to recording an NMR signal
encodes the signal with a phase which depends on the strength of the gradient,
the duration of the gradient and the position of the spins. Acquiring a series
of signals with different amounts of phase encoding, followed by a Fourier transform
allows the positional information to be decoded. Phase
rewinder gradient - a gradient which is applied after data collection and
has equal amplitude but opposite polarity to the normal phase encoding gradient.
Ensures that each interpulse interval produces the same net phase, allowing transverse
coherences to develop uniformly across the whole field of view. Phase
sensitive detector > Demodulator. Pixel
- acronym for picture element; the smallest discrete part of a digital image display. Planar spin
imaging - one particular technique of planar imaging that creates an NMR image
of a plane from one excitation sequence by selectively exciting a grid of points
within the plane and then applying a gradient magnetic field so that each point
has a different Larmor frequency. Fourier transformation of the FID can then be
used to separate the signals from each selected point and create the image. Planar imaging
- imaging technique in which the image of a plane is built up from signals received
from the whole plane. See also Sequential plane imaging. Point
imaging > Sequential point imaging. Point
scanning > Sequential point imaging. Polarisation
transfer - A technique for transferring magnetization from proton to strongly
coupled nuclei and hence improving the S/N ratio of the other (less sensitive)
nuclei. ppm
- part(s) per million. Precession
- comparatively slow gyration of the axis of a spinning body so as to trace
out a cone; caused by the application of a torque tending to change the direction
of the rotation axis, and continuously directed at right angles to the plane of
the torque. The magnetic moment of a nucleus with spin will experience such a
torque when inclined at an angle to the magnetic field, resulting in precession
at the Larmor frequency. A familiar example is the effect of gravity on the motion
of a spinning top or gyroscope. Precessional
frequency > Larmor frequency. Presaturation
- an excitation pulse followed by a spoiler gradient which is used to remove the
signal from a particular part of a spectrum or imaging volume. PRESS
- a technique for localized spectroscopy using a spin echo sequence. Probe
- the portion of an NMR spectrometer comprising the RF coils and some associated
electronics. The RF coils may consist of separate receiver and transmitter coils
in a crossed-coil configuration, or, alternatively, a single coil to perform both
functions. Program
> Software. Progressive
saturation > Saturation recovery. Projection
profile - spectrum of NMR signal whose frequency components are broadened
by a gradient magnetic field. In the simplest case (negligible line width, no
relaxation effects, and no effects of prior gradients), it corresponds to a one-dimensional
projection of the spin density along the direction of the gradient. Proton
density - (also: rho) the spin density of hydrogen-1 (1H) spins; represents
to a certain extent the water content of a sample or a tissue in proton magnetic
resonance imaging. Proton
density-weighted (rho-weighted) image - image whose contrast is mainly influenced
by proton density, but which also has T1, T2, and bulk flow contributions. Usually
used as a synonym of spin density, although this only applies for hydrogen. Better
term: intermediately (between T1 and T2) weighted. PS
> Partial saturation. PSIF
> CE-FLASH. Pulse,
90° (pi/2 pulse) - RF pulse designed to rotate the macroscopic magnetization
vector 90° in space as referred to the rotating frame of reference, usually
about an axis at right angles to the main magnetic field. If the spins are initially
aligned with the magnetic field, this pulse will produce transverse magnetization
and hence an FID. Pulse,
180° (pi pulse) - RF pulse designed to rotate the macroscopic magnetization
vector 180° in space as referred to the rotating frame of reference, usually
about an axis at right angles to the main magnetic field. If the spins are initially
aligned with the magnetic field, this pulse will produce inversion, for transverse
magnetization causes a phase change which leads to the generation of a spin echo. Pulse length
(width) - time duration of a pulse. For an RF pulse near the Larmor frequency,
the longer the pulse length, the greater the angle of rotation of the macroscopic
magnetization vector will be (greater than 180° can bring it back toward its
original orientation). Pulse
programmer - part of the spectrometer or interface that controls the timing,
duration, and amplitude of the pulses (RF or gradient). Pulse
sequences - set of RF (and/or gradient) magnetic field pulses and time spacing
between these pulses; used in conjunction with gradient magnetic fields and NMR
signal reception to produce NMR images. See also Interpulse times. Pulsed
gradients > Gradient pulse. Pulsed NMR/MR - NMR/MR techniques that use RF pulses and Fourier transformation of the NMR signal; have largely replaced the older continuous wave techniques. | |||||||||||||||||||||||||||||||||||||||||||||||||||||
|